01. Comparision test within the 0~5 NTU Range
Testing after spiking with Formazin short test(1h30)
9 levels of calibration (Lowest value : 0.35 NTU, Highest value : 5.1 NTU)
Comparison test results between The.Wave.Talk's module sensor and competitor's product divided into 9 concentration levels within the 0~5 NTU Range
The measurement values for the same sample between the two products show a high correlation with R²=0.9992, with no difference in the measurement values
02. Comparision test at low concentration(<1 NTU)
Blue line : The.Wave.Talk sensor / Gray line : Competitor's In-line turbidimeter / Orange dots : Competitor's Tabletop turbidimeter
The comparison test results between the The.Wave.Talk sensor and two competitor products within the low concentration range of 0~1 NTU, divided into four levels.
There is no difference in the measurement values between the The.Wave.Talk sensor and the competitor's Tabletop product.
Even in the low concentration range (< 1 NTU), there is no difference between the TWT sensor and the competitor's products.